cSQZ General
hudsonl - posted 19:54, Friday 31 May 2024 (11662)
Green Light Saturating SHG RFPD on Resonance
[Hudson, Jacques, Malo]

We observed that the compact squeezer's SHG PDH error signal leveled out at zero right at resonance when operating with high powers (see attached plot). We determined that this was due to green light produced on resonance leaking out through the input/output coupler and saturating the RFPD used for the SHG PDH lock. The error signal looked normal with the crystal adjusted away from the phase-matching temperature when no green light was produced. We fixed this problem by placing a 1064nm line filter in front of the RFPD to block the green light leaking out of the cavity (see attached plot for error signal with line filter in place). This seemed to substantially improve the intensity stability of the green beam produced by the SHG.
Images attached to this report