I replaced the new board in the UMZI 1km setup (always using the NKT laser).
Fig.1 shows the OLTF of the loop while the stage is not moving. The UGF is lower than before (12kHz), but this was the best configuration to keep the loop locked while moving.
Fig. 2 shows the error and correction signals acquired with the Moku:Go for the stage at rest and moving. The low frequency of the error signal has the same issue that needs to be fixed.
Fig.3 shows the time series acquired with the moku while the stage is moving and we notice that we don't have fringe wrapping anymore. This means that the change of the board was successful!
Fig. 4 shows the OLTF of the loop while the stage is not moving, with the new spool.
Fig. 5 compares the stationary data (stage at rest) acquired with the Moku and the SR785. In this case also the correction signal acquired with the Moku has a discrepancy with the data acquired with the SR785 below 700Hz.
Fig. 6 adds the data while the stage is moving. The correction signal is higher while moving between 200Hz-2kHz.
Fig. 7 compares the stationary cases with (SR785) and without box (Moku). It is evident that the situation with the box and the new servo reduces the correction signal between 70Hz - 3kHz.
Fig. 8 compares also the moving stage measurements with and withou box with the static ones.