Heisenberg IFO General
Malo - posted 12:46, Sunday 02 June 2024 (11656)
Homodyne detectors spectra
I removed the AC path on Jacques' board (board J) and consequently replaced the R34 500 Ohm by a 50 Ohm to avoid losing a factor 10 of voltage due to the 50 Ohm resistor inherent to the Agilent. I then added the sockets to both Eric's (board E) and Jacques' boards, so that we can use two HQE photodiodes for the moment and interchange them between the boards.

I then characterized both boards by making their transfer functions and noise spectra. The 1kHz-50MHz data will be useful for our current use of the HDs whereas the 1kHz-500MHz data gives us a broader view of the HDs, in particular above 100MHz the transfer functions fluctuate a lot but the noise is relatively stable. Most importantly, the DC port transfer function are relatively stable and flat from 1kHz to a few MHz, in both amplitude and phase, and the averaged noise spectra are relatively stationary.
DC OUT, RF1 and RF2 are the different outputs of the board. In particular, the AC signal comes out of DC OUT since we disconnected the AC paths.

-Noise spectra : they were done without illumination of the photodiodes.

We observe a peak at 13.6MHz coming from some ambiant noise in the lab that could come from the ground. Indeed when I disconnect the boards from the Agilent and let the SMA cable touch the optical table, the peak appears; whereas when I isolate the end of the cable it disappears. Moreover, on my setting the board doesn’t touch the table, so it seems to be transmitted by the power supplier: only when the power supplier is turned on, the peak appears. This peak was also visible in the other lab, in NW17.
We also observe some broadband noise around 22MHz in both boards, that should come from the electronics in the boards.
The noise around 10MHz and 15MHz must be a consequence of the RF ports, and it should be verified further that it won't affect the squeezing measure.

-Transfer functions : the signals were first done with a modulated laser illumination on one of the PDs. A comparison between the illumination of PD1 and PD2 shows that they are roughly equivalent, therefore we don't mention which one was illuminated on the later graphs.

The valleys in DC OUT are consistent with the peaks in  both RF ports.
A comparison of board E and J show that the functions are relatively similar, with only little difference in the amplitudes and a small frequency shift in the RF resonances.