epolini - posted 16:44, Wednesday 29 May 2024 (11653)
UMZI 1km NKT: measurements with moving stage

I made some measurements with the UMZI 1km NKT with the Moku:Go to characterize the effect of the moving stage (the horizontal Fuyu on the bench) with and without isolation spool and box. 

Comparison SR785 and Moku:Go spectra - static measurement

First of all, I compared the spectrum of acquired with the SR785 with the spectrum computed from the time series acquired with the Moku:Go data logger.

Fig. 1 shows the error and correction signals taken with both instruments using the plastic spool without the box.

We have the following parameters (as in the entry #11650):

I acquired 30s of data with the Moku sampling at 200kHz. There is a discrepancy not yet understood in the low frequency region of the error signal. 

Measurements with moving stage

1. Plastic spool - without the box

I moved the stage with the spool on top with the following settings:

X Tar Pos = 60

X Tar Speed = 0.7

X Acc / Dec = 4 

With the same parameters written before, I performed the measurements with the plastic spool. The comparison between error and correction signals for the moving and static cases is shown in Fig. 2. We can see that the curves relative to the moving case are too flat for frequencies < 1kHz. Moreover there is a broad peak at 2-5kHz.

I tried to change the loop parameters: increasing the VCO FM dev and reducing the servo gain. I took a new set of of measurements, both static and with the moving stage with the following parameters:

The new data are shown in Fig. 3. The issue is not solved.

Eric suggested that there was a fringe wrapping problem in this case that was also evident from the time series, as shown in Fig. 4. The error signal has a slope and the correction signal has issues following it as shown by the flat regions. We are not losing lock but the loop is re-locking really fast in another fringe, introducing the weird feature shown in the err/corr spectra.

The solution to this issue would be to change the servo board in order to be more aggressive in the suppression of the 2-5kHz oscillation. To do so we need to add one integrator at 5kHz and move the integrator at 482Hz up to 3kHz.

2. New spool inside the isolation box

I performed the measurements also with the new spool inside the box:

Static parameters: 

Moving parameters:

Fig. 5 shows the comparison between the static and the moving cases. The vibration and acoustic isolation provided by the new spool and the box avoids the fringe wrapping, as it can be seen also from the time series shown in Fig. 6. We notice the increase in noise in the moving case, as expected. The main contribution to the excess of noise while moving is due to the patch cords connected to the moving spool that is free to move.

Images attached to this report