cSQZ General
hudsonl - posted 20:16, Tuesday 28 May 2024 (11651)
CSQZ Squeezing Levels
After cleaning the optics the other week, I took a more careful squeezing and anti-squeezing spectrum today. The pump beam's power is still quite unstable, so it seems like it will be challenging to make a careful squeezing and anti-squeezing vs NLG plot until that is fixed. However, even with the unstable pump, we can start to investigate which parts of the setup are limiting our measured squeezing levels.

Currently, we measure 2.6 dB of squeezing and 6.73 dB of anti-squeezing measured between 55kHz and 70kHz, where we have a clean spectrum. I expect the squeezing measurement is more reliable here as it is less sensitive to pump power fluctuations. The last time we measured the homodyne, it had a visibility of 80%, and the OPA's escape efficiency is 0.95 giving about 40% total loss from these sources. As a result, I expect to see 4.0 dB of squeezing in the absence of phase noise, any additional sources of losses, and with the OPA approaching threshold. In the absence of phase noise and with the OPA near threshold, a maximum squeezing level of 2.6 dB is indicative of about 55% total loss, about 15% higher than that from sources identified so far. As this number assumes the OPA is very near threshold, which it is not, the unexplained losses are somewhat below 15%.

In conclusion, measurements of the compact squeezer so far are roughly compatible with squeezing levels limited by homodyne visibility.

We note that for given effective optical losses, L = (Power Out / Power In), the maximal observable squeezing level is -4.343*ln(L) [decibels]. This is plotted in the attached figure.