Heisenberg IFO General
jacques - posted 20:11, Wednesday 22 May 2024 (11648)
Building new homodyne detectors: fix to the old problem and new problems

[Eric, Jacques, Malo]

We had previously concluded that the AC OUT paths on both Eric's and my HD boards were not working properly. Eric debugged it by noticing that the PD sensors were shotcircuiting the auxiliary pins when they were fully shoved into the board. Thus the fix consisted in adding some electrical tape to insulate the aforementionned zone of the board and resoldering the PDs. 

But even with this fix, Eric's board's AC OUT was showing some increased electronic noise above a few hundreds of kHz, see pictures of the agilent. This was not noticed on the SR because of its limited bandwidth to 100kHz.

The DC path looked flat in terms of noise up to 10s of MHz, both on the Agilent and the SR, so the problem seems to come from the U8 op amp on the AC OUT (doing the inverting x10 gain). 

Thus we can still go on by desoldering the AC path's input capacitance and only using the DC path (which then also inherits the AC components), at the expense of losing the internal x10.

Next step:

1/ Do a TF measurement my shining modulated light on it.

2/ Debug my board by removing the AC path.