Faraday Isolator after NKT laser replaced

Today I replaced the HP fiber isolator (up to 2W) since it was damaged with a lower power one. This new one can take up to 300mW so we need to be careful not to increase the laser power above this threshold. 

I checked if the peak in the intensity noise (1kHz-tens kHz) was somehow caused or increased by such issue but it is not. The intensity peak is still there, as shown in the attached picture.