[Hudson, Jacques] We re-aligned the homodyne up to 80% visibility. After cleaning the optics, the homodyne got misaligned, so we took the opportunity to rework the homodyne beam path such that none of the optics for the Mach-Zehnder interferometer are on flip mounts. When that experiment is running, other mirrors, that would direct the beam directly to the homodyne will be flipped out of the beam path. We will now proceed to measure squeezing and anti-squeezing vs NLG and compute the total losses and phase noise to determine if these are consistent with the homodyne's imperfect visibility being the dominant loss mechanism.