Displaying report 1-1 of 1.
Reports until 19:52, Wednesday 08 May 2024
cSQZ General
hudsonl - posted 19:52, Wednesday 08 May 2024 (11630)
2dB Squeezing and 10dB Anti-Squeezing Measured with all Feedback Loops Closed
[Hudson, Jacques]

After replacing the custom Silicon PD measuring the CLF beam with a Thorlabs InGaAs detector, we were able to turn down the CLF power level considerably while retaining enough SNR on the CLF and LO loops to lock them.

With these loops locked, we measured about 2dB of squeezing and 10dB of anti-squeezing on the homodyne, verifying that all essential parts of the compact squeezer are working. Next, we will begin more careful diagnostics of the system and work to tune up the squeezing levels from 2dB to the 8-10dB range. The attached images show spectra with squeezing and anti-squeezing. Note that the y-axis scales are different in these images.
Images attached to this report