Reports until 18:10, Wednesday 08 May 2024
GRAVITES Optics 1550nm
epolini - posted 18:10, Wednesday 08 May 2024 (11628)
Noise Budget of the Balanced Mach Zehnder Interterferometer with 1km arm

I report here the noise budget of the 1km-BMZIs described in the entry #11626 using both the Thorlabs and the NKT lasers.

I also used the formula (23) of Chris' optica paper to express the maximum optical power as a function of the fiber length allowed by stimulated Brillouin scattering. In the equation appears the effective length (Leff) of the fiber. 

Fig. 1 shows Leff as a function of the real length for the Corning SM28-ULTRA fibers that we are currently using. 

Leff = [ 1 - exp(-a*L)]/ a

where a are the losses expressed in a linear scale. The SM28-ULTRA fibers have 0.18dB/km losses which translates into a linear scale as : a = 1 - 10^(-0.18/10) = 0.0406.

The Brillouin scattering threshold power is given by:

P_B = k * K * Aeff / (gB * Leff)

where k = 19 (numerical factor), K = 1 (mixing efficiency), Aeff = pi * rmf^2 = pi * (46 um)^2 (effective area), gB = 1.68e-11 m/W (Brillouin gain).

Fig. 2 show the maximum power due to SBS as a function of fiber length:

Psbs( L = 1km ) = 0.075W = 75 mW

Psbs( L = 40km ) = 0.0038 W = 3.8 mW

Noise budget BMZI 1km arms

I considered the model of the frequency noise of the lasers as shown in the entry #11616. I also included the rms of the correction and error signal data.

The input power P0 to the BMZI (before the first BS) = 3.2 mW.

1. NKT laser

Fig. 3 and Fig. 4 show the correction and error signal noise budgets respectively.

2. Thorlabs laser

Fig. 5 and Fig. 6 show the correction and error signal noise budgets respectively.

There is an underestimated source of noise in the thorlabs laser case. We just received the 100m and 200m spools, so I will build an UMZI with unbalance < 329m in order to measure the frequency noise.

Images attached to this report