epolini - posted 21:57, Monday 06 May 2024 - last comment - 13:25, Tuesday 07 May 2024(11626)
Balanced Mach Zehnder Interferometer Characterization: 1km (Thorlabs, NKT) and 40km (NKT) arms

Last Friday and today, I have characterized the balanced Mach Zehnder interferometer using both the Thorlabs and NKT lasers. Since the Thorlabs laser frequency noise is too high for an unbalance between the arms higher than 329 meters, we don't have at the moment a setup to measure it in this way. We ordered shorter spools but in the meantime I characterized the balanced MZI using both lasers.

1. BMZI characterization with 1km arms

It was possible to lock the loop using both the Thorlabs and the NKT lasers. I report the loop characterization for both.

1.1 Thorlas laser (BMZI 1km arms)

Sensor TF : 180 mV/rad

VCO FM dev : 6 kHz - VCO TF : 3.9 kHz/V

UGF : 32 kHz

Servo Gain : 33 dB

Fig. 1 shows the error and correction signal spectra calibrated in rad/rtHz.

Fig. 2 shows the OLTF of the loop.

Fig. 3 shows the derived TF of the servo board.

1.2 NKT laser (BMZI 1km arms)

I increased the laser power in order to have the same error signal fringes amplitude.

Sensor TF : 175 mV/rad

VCO FM dev : 6 kHz - VCO TF : 3.9 kHz/V

UGF : 31 kHz

Servo Gain : 33 dB

Fig. 4 shows the error and correction signal spectra calibrated in rad/rtHz.

Fig. 5 shows the OLTF of the loop.

Fig. 6 shows the derived TF of the servo board.

2. BMZI characterization with 40km arms

It was possible to lock the loop ONLY with the NKT lasers. The most probable reason is that the unbalance between the two 40km spools is of the order of hubndreds of meters and this prevents the loop to be locked due to laser frequency noise. I characterized the BMZI loop with the NKT in two different configurations: same settings as before that gives gain peaking and lower gain on the servo to avoid gain peaking.

2.1 NKT laser (BMZI 40km arms) - same settings as before

The error signal fringes have a smaller amplitude.

Sensor TF : 154 mV/rad

VCO FM dev : 6 kHz - VCO TF : 3.9 kHz/V

UGF : 30 kHz

Servo Gain : 34 dB

Fig. 7 shows the error and correction signal spectra calibrated in rad/rtHz.

Fig. 8 shows the OLTF of the loop.

Fig. 9 shows the derived TF of the servo board.

2.2 NKT laser (BMZI 40km arms) - lower gain

Sensor TF : 154 mV/rad

VCO FM dev : 6 kHz - VCO TF : 3.9 kHz/V

UGF : 1.56 kHz

Servo Gain : 7 dB

Fig. 10 shows the error and correction signal spectra calibrated in rad/rtHz.

Fig. 11 shows the OLTF of the loop.

Fig. 12 shows the derived TF of the servo board.

 

Next steps:

We need to measure the unbalance between the two long spools in the arm. Eric suggested to modulate the frequency of the laser and then measure the phase at the output of the interferometer. I ordered some spools (100 meters, 200 meters and 1 km) so that we can balance the arms and lock the 40km-BMZI also with the thorlabs laser. 

Images attached to this report
Comments related to this report
epolini - 13:25, Tuesday 07 May 2024 (11627)

As a comparison, I plotted on the same plot the error and correction signals calibrated in rad/rtHz of the BMZI 1km with NKT and thorlabs lasers. The loop configuration is comparable. We notice an excess in noise at frequencies > few kHz for the measurements done with the thorlabs laser. The plot is attached.