Displaying report 1-1 of 1.
Reports until 19:42, Wednesday 08 May 2024
cSQZ General
hudsonl - posted 19:42, Wednesday 08 May 2024 (11625)
CLF RFPD Path Modifications
I added a steering mirror after the beam splitter in the CLF path acting as a circulator. The steering mirror allowed me to direct the CLF beam reflected from the OPO off of the compact squeezer board and onto the table, where I set up a Thorlabs high gain, high bandwidth InGaAs PDA10CF. See attached image.

I anticipated that this PD would have better performance than the custom PD since it uses an InGaAs diode instead of a silicon diode. However, I anticipated this improvement would be roughly a factor of 4, accounting for the quantum efficiency of InGaAs (0.7 A/W at 1064 nm) compared to silicon (0.15 A/W at 1064 nm). However, the Thorlabs detector had a noise performance almost 20 dB better than the custom Silicon PD. 

Switching out these PDs allowed us to significantly reduce the CLF power while retaining reasonably high SNR on the CLF and LO feedback loops. 

The attached file shows the beam path with the new Thorlabs PD.
Non-image files attached to this report
Displaying report 1-1 of 1.