Heisenberg IFO General
jacques - posted 20:29, Friday 03 May 2024 - last comment - 16:17, Sunday 05 May 2024(11621)
building the new HDs: origin of the 43kHz oscillation & other preliminary tests

I debugged by HD board:

- At the start of my inquiries, spontaneously, the DC power supply would show:

1/ That the +18V is drawing 0.16A while the -18V is drawing 0. The cause of this is not certain but I am suspecting the power cable to play a role in it

2/ That both +-18V are drawing 0.3A, which is much higher than the expected ~0.1A in normal regime. The origin of that is probably a bad soldering of one of the components, leading to either bad contact or shorting of a component. The way to solve this is to remove the most suspect components until the current drawn is back to a reasonable value.

- Symptom 2/ was accompanied by the U4 and U3 regulators (maintaining +-5V for the rest of the board) which became exceedingly hot. This meant that the component which was drawing all this current was one the OpAmp powered at +-5V. It turned out that this was U11, as removing it made the DC supply go down to <0.1A. 

- The 43kHz oscillation comes from the RF2 path (notch at 10MHz), as removing either the U11 OpAmp or the L11 inductor made these parasitic oscillations disappear in all of the other outputs. It is however important to note that the C49 capacitor was always removed, so the signal path is cut. This means that somehow the RF2 path picks up noise from the environment and oscillates at its frequency? Unless there is some bad soldering on the RF2 path...

- U9 can be resoldered using Slawek's heat gun (but make sure that the chip is mainted in place using a tweezer). Careful examination of the pins's sides should determine whether U9 has been soldered properly or not.

- I then did some quick tests of the outputs: I tested the PD1 and PD2 paths, simply by shining a lamp on both photodiodes at the same time and looking at the outputs on a DC scope. I do get a non-zero DC output on the scope (see attached): a periodic signal at the grid frequency of 120Hz. This tells me that the paths to the PD output are working. It is important to notice that neither signal go to zero when one I block one of the PDs from the incident light; I would have expected the corresponding PD output voltage to go to 0 but I could misunderstand these outputs.

- I also get a signal on the DC OUT path, which goes from -5V to +5V depending on the orientation of the board wrt the light source.

- The AC path also gives a non-zero output when I shine some light on the PDs

- RF1 gives a very strong and stable oscillation at 16.66MHz (see picture), even when there is no light on the PDs, which is very worrying.

Images attached to this report
Comments related to this report
jacques - 16:17, Sunday 05 May 2024 (11624)

I realize that R62 (tranimpedance) was not soldered in during these tests.

Below is a noise plot obtained at the AC OUT when all the components are stuffed except U10 and U11, there is 25dB of attenuation on the agilent input.

I don't see any noise on AC OUT of Eric's board so there is also something broken there.

Images attached to this comment