GRAVITES Optics 1550nm
epolini - posted 21:16, Wednesday 01 May 2024 (11616)
Maximum unbalance between the interferometer arms before being limited by laser frequency noise

Since one possible issue with the UMZI not locking could have been an excess in laser frequency noise at the input of the Mach Zehnder (and Michelson for the Thorlabs as well), we calculated the maximum length unbalance to keep the phase at the output below 1 radian.

I did that for both the NKT HP X15 and the Thorlabs ULN15TK. I fitted the frequency noise taken from the datasheets. I computed the rms for the input frequency noise. I used the formula to convert the input frequency noise into output phase noise used in Chris' paper:

S_phi(omega) = (2 pi tau)^2 * S_freq(omega) -> A_phi(omega) = (2 pi tau) * A_freq(omega)

where tau = n*DL / c.

We set A_rad_rms = 1 rad.

1. Thorlabs ULN15TK maximum arm unbalance

Fig. 1 shows the laser frequency noise data, fit and rms. I considered A_rms_freq = 1e5 Hz.

DL = A_rms_phi * c / (2 * pi * n * A_rms_freq) ~ 329m

This means that with the arms unbalanced by 1km, as it is in the current setups, we will have fringe wrapping at the output. 

2. NKT X15 maximum arm unbalance

Fig. 2 shows the laser frequency noise data, fit and rms. I considered A_rms_freq = 5982 Hz.

DL = A_rms_phi * c / (2 * pi * n * A_rms_freq) ~ 5505m

This confirms that having the arms unbalanced by 1km should be fine for this laser.