epolini - posted 20:57, Wednesday 01 May 2024 (11615)
Lessons learned debugging the UMZI - 1km loop

The UMZI was not locking in the past days even with the NKT laser. While debugging, we measured the servo board and adjust the settings on the marconi. 

1. Eric's servo board transfer function from OLTF

The first thing we tried was to match all the parameters of the loop we had once we were able to lock it. Sensor TF ~ few tens mV/rad, LO 80MHz and 7dBm, VCO 80MHz and -5dB (sent to a HP amplifier), PI corner at 480Hz (turning integrator 1 ON on the board) and then adjust the gain accordingly (there is no indicator, so it needs to be measured). To derive the gain used when locked, I did:

TF_servo = OLTF / (TF_sensor * TF_Plant * TF_VCO) 

Using the OLTF, TF_sensor, TF_Plant, TF_VCO used in the entry 11479.

freq_27 = np.array(freq_27)
mag_27 = np.array(mag_27)
phase_28 = np.array(phase_28)
OLTF_complex = 10**(mag_27/20)*np.exp(1j*np.pi*phase_28/180)
c = 3e8
L = 1e3
n = 1.44
tau = n * L / c
P = np.exp(-1j*2*np.pi*freq_27*tau)
A = 6.5e3 / (-1j*freq_27)
S = 25e-3
F = OLTF_complex / (P*A*S)
F = np.array(F)

magnitudes = np.abs(F)
phases = [cmath.phase(element) for element in F]
phases_degrees = [phase * 180 / cmath.pi for phase in phases]

mag_db = 20*np.log10(magnitudes)

plt.figure(figsize=(9, 6))
plt.subplot(2, 1, 1)  
plt.semilogx(freq_27, mag_db, label = 'Servo Board Mag')
plt.ylabel('Magnitude [dB]')
plt.xlabel('Frequency [Hz]')
plt.xlim([100, 1e5])
plt.legend()
plt.grid()


plt.subplot(2, 1, 2) 
plt.semilogx(freq_28, phases_degrees, label='Servo Board Phase')
plt.ylabel('Phase [deg]')
plt.xlabel('Frequency [Hz]')
plt.xlim([100, 1e5])
plt.legend()
plt.grid()

The plot is shown in Fig. 1. The gain to set on the board is the one at high frequency in the TF, meaning 40dB.

2. Set 40dB gain on the servo board by looking at its TF on the SR785

To set the gain correctly, we tune it looking at the servo's TF. We send the swept sine to the input of the servo board, i.e. where we usually send the error signal. Then we take the output at the output of the board, i.e. where we usually take the correction signal. We put the signals on the SR785 and we see the gain range of the board by tuning the gain. The range is 6dB to 40dB. 

NB. be careful not to saturate the board by sending a large amplitude swept sine. 500mV amplitude was making saturating the board and I was reading a gain range of -28dB to 5dB. A good amplitude is 50mV.

We set the gain back to 40dB.

3. Coupling settings on the Marconi

I did not notice that the settings on the Marconi changed. The FM modulation was set to Ext AC mod coupled instead of Ext DC mod coupled. By checking the output of the marconi with the frequency counter it was obvious that the correction signal of the loop was saturating but the frequency was not spanning the VCO range.

 

Images attached to this report