GRAVITES Optics 1550nm
epolini - posted 20:21, Wednesday 01 May 2024 (11614)
Thorlabs laser power measurement after turning ON

Today I acquired again power data from a moment the laser was off and for 4 hours after turning it on, to understand the time we need to wait for the power to stabilize. 

Fig. 1 shows the full data points at the top and then I zoomed into the data when the power was on. There is a power drift at the beginning and then it fully stabilize after ~1h30. 

I inserted a BS 95/5 after the FI following the laser. The input of the BS is port 1 and the outputs are:

Port 2 (5%) : P = 4.5mW

Port 3 (95%) : P = 79mW.

Port 3 (95%) goes to the full setup, while the 5% pick off goes to another 99/1 BS. The 1% of this BS goes to the PDA20CS photodiode connected to the Moku:Go.

The power measured with the powermeter is 35uW while the recorded value is a bit lower. This can be due to the not perfect value of the transimpedance gain reported on the datasheet or the clipping due the beam divergence in the photodiode (since the sensitive aperture of the powermeter is substantially larger).