Reports until 17:07, Friday 26 April 2024
cSQZ General
hudsonl - posted 17:07, Friday 26 April 2024 (11611)
Dichroics Added to Path from SHG to OPO
I added a pair of dichroics on the path from the SHG to OPO on the compact squeezer to prevent IR light from leaking from the SHG to the OPO. Each of these dichroics is specced at <10^-4 IR transmission. 

Previously, we had about 50mW of IR into the SHG, with a finesse of 150, so we expect about 7.5 W of circulating IR power in the SHG (ignoring pump depletion for simplicity in the order-of-magnitude calculation). We then expect about 750 uW transmitted through the SHG's dichroic and 75 nW sent into the OPO's input/output coupler after the single dichroic on this beam path. This ignores losses from the commercial 532nm optics on the PDH board, which will probably attenuate the IR beam by a few orders of magnitude so we're sending closer to 1 nW into the OPO. 

After adding a pair of dichroics, we should now be sending at most 1 fW of IR power into the OPO's input/output coupler and realistically are likely sending closer to 10 aW into the OPO.

After adding the dichroics, we re-aligned the OPO but did not redo the mode matching. On resonance, we now have about 150 mV on the 532 transmission PD compared to 500 mV previously, indicating that the mode matching is substantially worse than it was previously. This is not surprising given that the beam now passes through a pair of dichroics which we expect to shift the waist position. The dichroics are 6mm thick with an index of about 1.5 and are passed through at 45 degrees, so we expect the beam's focus to shift by about 2*(1.5-1)*6*1.41 = 8.5 mm.

The thermal effects in the OPO also appeared to be worse with this configuration, indicating that improving the mode matching may also improve the OPO's thermal stability.