[Eleonora, Eric]
We spliced the 40-km spool with our splicer in green lab following the same procedure of last time and everything went smoothly.
Fig. 1 shows the results on the splicer for the external end of the spool - loss = 0.01dB due to axis offset
Fig. 2 shows the results on the splicer for the internal end of the spool - loss = 0.02dB due to axis offset and core bending
This time we also installed the strinking tubes bought on amazon that Gabor suggested me to get, as shown in Fig. 3.
In order to have both ends of the spliced spool exiting from the top, I copied how Gabor did with the other spool, as shown in Fig. 4.
I measured the losses of the spool (using the NKT laser):
Input power = 113 uW
Output power = 20.5 uW
Loss = 82%
Loss_dB = 10*log10(113/20.5) = 7.4dB.
This loss level corresponds to 0.185 dB/km which is in agreement with what we expect, meaning that the splicing was successful.