I noticed that the new thorlabs laser power fluctuates over time. I noticed it yesterday and I checked in different locations: right after the laser, after the isolator, after couple of BSs. I left the laser on the whole night and this morning the power was 20% higher than yesterday evening. We have an appreciable 20% decrease/increase also on a scale of tens of minutes. I queried the laser status and temperature with the commands shown on the manual using pyserial and the answers are not in agreement with what stated there. I contacted the thorlabs technical support to understand how to stabilize internally the laser temperature and (hopefully) stabilize the power.
I saved the time series data using the Moku:Go, sampling rate 10Sa/s. I acquired 4 sets of 5 minutes. I used the PDA20CS with gain 0dB which corresponds to a transimpedance gain (TG) of 1.51e3 V/A.
I attach the data expressed in Volts and in Watts.
Power[W] = Vout / (R(lambda) * TG)
The laser current is stable at 0.55 A.
The laser temperature is stable at 25 Clesius.
The fbg temperature is stable at 28 Celsius.
Today we measured the power after placing a 95/5 BS and a PBS after the FI. Attached the power at the output of the two PBS's ports. Pol 2 fluctuates more than pol 1.
I acquired data last Friday during the day (~2h30) and in the evening (~12h), after noticing that the saturation of the PDA20CS was ~2V for 0dB gain setting which corresponds to a maximum input power of 1.4mW.
Fig. 1 and Fig. show the time series of the power registered with the PDA20CS. The power is not fluctuating as before, but I am keeping the laser on in order to have the laser more stable. The power flucuates by less than 5%.