Yesterday, Jacques, Eric and I noticed that we were unable to excite only the vertically polarized cavity modes by adjusting the half waveplate between the 532nm Faraday isolator and the OPO. We verified that all of the optical elements are working properly. The problem seems to be that there are several mirrors between the Faraday and the waveplate, which likely have different reflection phases for vertical and horizontal polarizations. As a result, the light is elliptically polarized by the time it reaches the half-waveplate. Given the setup's space constraints, the best solution seems to be to place a 532nm quarter waveplate next to the half waveplate on the beam path from the SHG to OPO.