Reports until 20:17, Saturday 13 April 2024
GRAVITES Optics 1550nm
epolini - posted 20:17, Saturday 13 April 2024 (11591)
Measurement UMZI 1km on Fuyu linear stage pt.2

1. Low pass filters at the Moku's inputs

I acquired new data with the Moku:Go after inserting two low pass filter, with frequency adapted for the sampling frequency, at the Moku's inputs to avoid the aliasing. The maximum time serie that the oscilloscope can acquire is 100-seconds long and the sampling frequency is 100Hz (ts = 10ms). The low pass filter has been chosen to have a fc ~ 25Hz in order to filter all the unwanted folded noises. In the experimental setup, the PD should be connected to the LPF and then the signal propagates until the Moku:Go. The output resistance of the PD 1811 is 33Ohm while the Moku load impedence is 'high'. I chose a 100kOhm resistor and a 67nF capacitor. The obtained fc = 24Hz. I soldered two LPF inside small boxes, as shown in Fig. 1, and connected them to the Moku, as shown in Fig. 2. 

2. Data acquired with the Moku:Go

I used the oscilloscope on the Moku:Go. Channel 1 is connected to the error signal and channel 2 to the correction signal. The loop was locked for the entire time of the measurements. I exported the data as file.csv, settings.txt and screenshot.png. Then I imported the data in python and plotted the time series with the following script:

t = []
err_V = []
corr_V = []

# Path to your CSV file
csv_file_path = 'MokuOscilloscopeData_20240413_185302_Traces.csv'

# Read data from CSV and store in arrays
with open(csv_file_path, 'r') as file:
    csv_reader = csv.reader(file)
    for row in csv_reader:
        if row[0].startswith('%'):  # Skip lines starting with '%'
            continue
        t.append(float(row[0]))   
        err_V.append(float(row[1]))  
        corr_V.append(float(row[2]))  


t_shift = [time - t[0] for time in t]    
plt.plot(t_shift, corr_V, label = 'corr signal')
plt.plot(t_shift , err_V, label = 'err signal')
plt.legend()
plt.ylabel('Amplitude [V]')
plt.xlabel('Time [s]')
plt.grid()        

I show the time series data for three different configurations: the stage is not moving, the stage is moving towards the computer, the stage is moving towards the wall.

2.1 Stage not moving

Fig. 3 show the data without the 24Hz-LPF in front of the Moku's input. Fig.4 shows the correct time series data, after the insertion of the LPF. The effect of the filter is clear.

2.2 Stage moving towards the computer

I used the same setting mentioned in the entry 11588, i.e. position +60, acc/dec 5, speed 1, Rel. The filtered data are given shown in Fig. 5. I started moving the stage after ~10 seconds and it can be seen that the correction signal increases. The stage finished its ride after ~60 seconds and it can be seen that the correction signal desreases.

2.3 Stage moving towards the wall

The only different setting here is the position -60. The filtered time series is shown in Fig. 6. The noise of the stage moving in the two directions is a bit different and I am wondering if this effect is also shown in the time data. Also here it is clear when the stage was moving looking at the correction signal.

Next steps: to perform and tune the PSD of these data and understand the figure of merit to look at to determine the performance of the different stages (high load Fuyu and rotation actuator Zaber).

Images attached to this report