In the past days I arranged the optical setup to be able to displace the 1-km spool on the stage without encountering obstacles. I connected the Moku:Go inputs to the error and correction signals closed loop of the UMZI setup. The powers coming from the two arms and recombining in the second 50/50 BS of the UMZI, are 122uW (I tuned the VOA). The calibration of the error signal is 67 mV/rad. The loop was closed as done last time.
1. Stabilization of spool and stage
At first, since while the stage was moving the spool was not stable, I screwed a post into the small platform hosting the spool on the stage and I added some ring spacers to fit the height. The I insert the black piece shown in Fig.1 to fix the spool to the platform. I also fixed the stage to the optical bench with three clamps (see Fig. 2). NB. if the moving platform eccesses the range, the stage is over vibrating and this makes the stage itself displacing, despite the clamps.
2. Chose the speed and acceleration of the moving platform
I tried different options for the speed and accelation. I was checking the error and correction signals on the oscilloscope, both on the analog oscilloscope and the Moku:Go oscilloscope, which I used to acquire time series data.
How to set the controller interface:
Select the single-JOG interface.
X Tar, X Tar Speed, X Tar Acc, X Tar Dec are respectively the target position, speed, acceleration and deceleration. Then one can chose between Abs (absolute), Rel (relative) and Stop. Abs refers to the absolute value for the position and after testing this feature, I realized that there is an hysteresis and then it is better to use the relative option. I move the stage manually to one and and then I understood that to cover almost all the linear range I should indicate +/-60 in the X Tar Pos. The sign + refers to the direction towards the computer, the sign - to the direction towards the wall. To avoid being too violent on the spool and to unlock the system, I set X Tar Speed to 1 (that should correspond to 1mm/s according to the manual) and X Tar Acc/Dec = 5 (see Fig. 3). I measured the distance and the time relative to these settings and I got: 150mm in 60 seconds that corresponds to 2.5mm/s (speed = 1). Then I changed the X Tar Speed to 2 and I got: 150mm in 30 seconds that corresponds to 5mm/s.
3. Measurements of the error and correction signals and polarization changes during the stage motion
I installed two PDA20CS PDs at the outputs of the PBS after the polarization controller that was ON during the measurements. Since the Moku has only two input channels, I monitored the polarization time series only with a traditional oscilloscope. In Fig. 4 are shown: error signal closed loop (yellow), correction signal (blue), signals in output of the PBS after the polarization controller (purple and green) while the stage with on top the spool is moving with the settings already discussed. We can see that the purple signal is changing over time, but then it goes back to its average value. The correction signal has some spikes but overall the system seems locked for the whole measurement time.
4. Check OLTF of the UMZI before and after the measurements with moving stage
I measured the OLTF of the loop injecting a swept sine of 300mV amplitude, both before and after all the measurements with the stage and it did not change as shown in Fig. 5.
5. Power check after the measurements
The powers recombining at the second 50/50 BS are 117uW (reference arm) and 113uW (spool arm). The power decresed but it did not end up to be too different between the two arms.
Next step:
I acquired the time series with the Moku:Go and I will show the spectra while the stage is at rest and when it moves.
NB. the powers in Watt need to be divided by a factor 10 here since there was an attenuation on the powermeter but no ND filter installed during these measurements.The powers recombining at the second BS are of the order of 10 uW and not 100 uW.