jacques - posted 19:37, Wednesday 03 April 2024 (11583)
Long term stability of the PLL + first implementation of the notch filter

1/ I switched the fast and slow loop on at 12:15 and came back at 14:15, with the loop still on. The temperature loop is able to keep the fast correction voltage in [0.4V, 0.6V] for timescales >2h.

Conclusion: the temperature control of the PLL works (provided that the temperature of the lab stays stable)

However, the noise of the error signal changes by quite a lot over time, without any apparent change to the loop nor to the setup. This is more pronounced at the end of the afternoon, when I could even clearly see the beatnote (free running without any loop closed) quite regularly jumping by >10MHz in a matter of a few seconds. I pin this on the bad weather which could have changed the conditions of the lab in a timescale smaller than the HVAC and thus destabilized the internal temperature loops of the lasers. The solution to this could be to feed the error signal of the fast loop to the slow one.

2/ I added the IIR filter to the PID block. It is a simple notch filter at 240kHz and span 100kHz. See figure where I took a OLTF of the loop, with and without the IIR. The IIR module does work but the notch is too narrow for now. A solution could be to add another notch close to the current one.

Note that the parameters I had to feed to the IIR are: 

    pole_2 = 1/2*(-f_span_IIR +1j*np.sqrt(np.abs(f_span_IIR**2 - 4*f_center_IIR**2)))
    zeros = [1j*f_center_IIR+300]
    poles = [pole_2]