I recovered the PLL with the new RP and took a spectrum.
The parameters which ensured good stability today were: P=0.01, I=20, no input filter.
Below is the phase noise plot. I recover the ~30mrad I had before. Not great, not terrible.
I also took some transfer functions using both the SR785 modulating on the Moglabs PM port, and using the RP's NA module. Analysis will follow tomorrow, although it seems that the piezo resonance at ~200kHz may add noise. So a spectrum using the Agilent will be needed.