Heisenberg IFO General
jacques - posted 16:21, Monday 01 April 2024 (11578)
Transfer function of the PLL RP

I measured the transfer function of the RP's PI block using the SR785, for two sets of PI values.

- Figure 1 shows the transfer function with the following parameters. I need to put an input filter in the RP's PID block in order for the signal not to saturate because of the low frequencies integration. The command is   pid1_pll.inputfilter=[-f_HP] where f_HP is the corner frequency of the filter in Hz. It is also important to reset the integrator value to 0 each time so as to discharge the accumulated drift from low frequencies. Maybe this was the problem I encountered on Friday (entry 11572). Parameters: SR785 source at 10mV, P = 2, I = 5000, f_HP = 200.

- Figure 2 shows the transfer function with SR785 source at 200mV, P= 0.09, I= 45, f_HP = 1.

They are both compatible with the theoretical model (with more deviation in the first case).

Conclusion: The RP seems to be working well between 10Hz and 100kHz. To understand if there is phase loss at higher frequencies, I need to switch to the Agilent.