Reports until 11:13, Friday 29 March 2024
Heisenberg IFO General
jacques - posted 11:13, Friday 29 March 2024 - last comment - 11:59, Friday 29 March 2024(11568)
Transfer function of the PZT driver

Here is the bode plot of the frequency response of the PZT driver, using the SR785 with source input of 50mV. It seems to operate as intended and does not explain the peak at 14kHz that was spoiling the phase noise (see entry 11535).

The procedure to measure a frequency response can be found in page 1-11 of the SR785 manual, or in Nutsinee's youtube channel [https://www.youtube.com/watch?v=4cS-ya6siRk]. Note that although, in dual channels mode, the bottom plot is called "B Live", it is still measuring the signals from ports A of both channels in "Frequency response" mode. Also note that you need to "print screen" both plots separately to get the amplitude and phase of the TF.

Note: to switch the frequencies to log scale, push the "Freq" button and select "Type" to be "log".

Images attached to this report
Non-image files attached to this report
Comments related to this report
jacques - 11:59, Friday 29 March 2024 (11569)

This can be modelled by an all-pass filter with different gains at high and low frequencies (see attached for the parameter specifications). Below is the comparison between the measured bode plot and the model. This matches very well so the behavior of the PZT driver is well understood.

 

Images attached to this comment
Non-image files attached to this comment