epolini - posted 12:36, Thursday 28 March 2024 (11565)
Thorlabs homodyne PDA10BI adapted seems not working

After adapting the setup to be able to install the homodyne detector, I tested the PDA10BI adapted as explained in the entry #11561. 56uW are arriving on both PDs. The new fiber port is on the input + photodiode. I checked on the oscilloscope the channels monitor + (yellow), monitor - (blue), RF output (purple) shown in Fig. 1. Monitor + is not seeing all the power, monitor - sees 60mV. The RF signal has an offset of -128mV. Fig. 2 shows also the feature of monitor + vs monitor - and it seems more noisy. The sensive area is really small so I doubt that the problem is that the light is not reaching it. The fact that the fiber port was missing suggests that maybe it was not working in the past and people removed it for diagnostic purposes.