Reports until 12:24, Thursday 28 March 2024
GRAVITES Optics 1550nm
epolini - posted 12:24, Thursday 28 March 2024 (11564)
UMZI with 1km spool full characterization with few changes

Yesterday I restored the original setup of the UMZI with 1km setup needed for the test of the stage. I adapted it on order to test the thorlabs homodyne that we recently interfaced with fiber ports and eventually a low noise homodyne (fempto or aLIGO-like). 

I moved the 90/10 BS from the end (right before the 1811) to the beginning (right after the FI). I added a VOA in the short arm in order to balance the power inside the arms. I used the Marconi we moved back from NW22 as LO. I re-connected the polarization controller and enabled it. Fig 1 shows the power budget of the current setup. 

Then I closed the loop using the 1811 PD, before testing the homodyne. The DC power arriving on the 1811 is 112mV and the RF power is 200mV pk-pk (~-10dBm). I mixed this signal with 7dBm LO and the error signal after a 1.9MHz LPF was 72mVpk-pk. This translate in a calibration factor of 36mV / rad.

The error and correction signals are shown in Fig. 2. The OLTF is shown in Fig. 3 (NB. I added 20dB to the mag as required by the design of the board). I sent a sweep sine of 300mV amplitude on the SR785. The UGF is ~20kHz. 