Reports until 10:31, Wednesday 27 March 2024
cSQZ General
hudsonl - posted 10:31, Wednesday 27 March 2024 (11562)
CSQZ OPO HR Mirror Reflectivity
We observed that the OPO's transmission of its CLF beam is about 0.7%, whereas its expected transmission is 1.7%. At the same time, the cavity's measured IR finesse of 54 agrees well with the expected finesse of 51 and indicates that there are no large excess losses. The OPO's mirrors should be coated for 89% reflectivity on the input/output mirror, and 99.95 +- 0.02% on the HR mirror (see attached quotation from LaserOptik). Some of the disagreement in measured power likely comes from mode mismatch, but since the cavity scan looks fairly clean (see attached image), I do not believe this can fully explain the factor of 3 difference between the expected and measured OPO transmission. 

I believe the measured transmission is likely lower than anticipated due to the HR mirror having a slightly larger reflectivity than 99.95%. For instance, if the mirror's reflectivity is really 99.98%, the expected finesse and transmission numbers agree exactly with the measured values. LaserOptik measured the reflectivity at 99.95% (see attached pdf), but this was done on a flat witness sample instead of the highly curved cavity mirror.

Given these measurements, my baseline assumption going forwards is that the OPO's HR mirror has a reflectivity of about 99.98% for IR.

Note: on the mode scan, channel 1 is the cavity length and channel 2 is the transmitted CLF power. In the LaserOptik_Final document, the OPO's HR mirror is "M4".
Images attached to this report
Non-image files attached to this report