The attached image shows an updated power budget for the Heisenberg interferometer, which should allow us to achieve about 135mW of pump power into the SHG if necessary, compared with about 44 mW in the current setup. It will also allow us to pick off all of the local oscillator beams from the Lightwave electronics laser with several mW in each beam. The main improvements will come in two places (1) swapping a fiber isolator for a free-space Faraday isolator with a measured 7% loss and (2) using a high-power fiber beam splitter for the initial 90/10 power splitter, which will allow us to improve the mode matching into the fiber. The mode-matching will need to be re-done after adding the Faraday anyways, so this is a convenient time to redo the mode matching. Moving the PLL pickoff to after the 90/10 beam splitter will also reduce losses on the SHG pump path, but by less than 10%, so this is a more minor improvement. As noted in the figure, the power budget assumes measured losses where available, and the worst-case scenario specification from AFW otherwise. AFW quotes a max excess loss of 1dB per splitter, corresponding to 7% loss per splitter. I measured the Faraday to have 93% transmission, in line with the manufacturer's spec of less than 8% loss. The assumed transmission through the fiber port is 50%, which is the current value, although this may be improved by purchasing fiber couplers with a longer Rayleigh range in the future.