Reports until 17:27, Wednesday 20 March 2024
cSQZ General
hudsonl - posted 17:27, Wednesday 20 March 2024 - last comment - 17:31, Wednesday 20 March 2024(11552)
Compact Squeezer Homodyne Improvements and Alignment
I modified the compact squeezer's homodyne in several ways. First, I removed the 50mm focal length lenses that had been used in the setup as they were not super-polished and replaced them with a pair of super-polished lenses of the same focal length. (2) I mounted all of the optics on 1 inch diameter, 3 inch tall posts, with the exception of the lenses and the beam dump for the rejected power. This will hopefully improve the detector's mechanical stability. The attached file shows the current homodyne.

After making these modifications, I re-aligned the homodyne. It is currently shot noise limited above about 1kHz, and the shot noise clears the detector's dark noise by over 20 dB. The full spectra are shown in the attached plot. I also ensured that there is a power plateau as the steering mirrors after the beam splitter are scanned vertically or horizontally, so the beams should be hitting the centers of the diodes and should not be clipped. i.e. the beam spot size on the diodes should be smaller than the diodes themselves.
Images attached to this report
Comments related to this report
hudsonl - 17:31, Wednesday 20 March 2024 (11553)
The LO power was 2.0 mW while measuring the shot noise spectrum.