Yesterday I tried to investigate better the issue I encountered last week of not being able to stably lock the Michelson. In mid-February I was able to lock it while recently I was not able to keep the lock.
The loop seemed not to correct entirely for the noise. Fig. 1 shows the OLTF which is losing coherence at high frequencies and Fig. 2 shows the error and correction spectra. The loop configuration is not ideal and Ben suggested to introduce the LB1005 after using the SR560 as pre-amplifier.
We set the DC gain of the SR560 two steps below the overload condition, around 50/100. Then we set to zero the input channel disconnecting the LB from the system. After playing with the servo settings we managed to keep the loop locked for most of the time, but still with some spikes as shown in Fig. 3. It shows the error signal (input, yellow), correction signal (output, purple) and the error monitor (blue) on the oscilloscope. We can see that the system is locked for most of the time if the environment is quite (no talking, no tapping, no walking etc). The spikes identify that the loop loses lock and it is therefore not stable enough. We tuned the parameters of the LB in order to get the system stable for the longer possible time.
Some reasonable parameters:
I was not able to acquire an OLTF because the system was losing lock. I acquired some spectra even if they are not perfect due to the lose of lock, which is partially filtered out by avaraging (Fig. 4).
We cannot do better with the current isolation of the Michelson interferometer, so we need the enclosure we have been working on with Medha. It is possible to lock the interferometer so I guess that RP can be currently tested.