cSQZ General
hudsonl - posted 11:42, Friday 08 March 2024 (11546)
OPO Thermal Control Electrical Insulation
The compact squeezer's ovens have problems if the 'ground' of the coax cables connecting the thermistor and thermo-electric controller (Peltier) to the controls box come into contact with electrical ground. In these circuits, a voltage is applied across the thermistor, and a current across the thermo-electric controller and in both cases the 'ground' pin is not at electrical ground. For the SHG, there was no grounding issue as the connectors between the jumper cable attached to the thermistor and thermo-electric controller and BNC cables are dangling from the table. For the OPO, this problem was fixed by wrapping the connection in electrical tape, as shown in the attached image.
Images attached to this report