cSQZ General
jacques - posted 19:57, Wednesday 06 March 2024 - last comment - 16:21, Friday 08 March 2024(11544)
second HD shot noise limited

I put the BS of the second HD on a newport mount. It is mounted such that the HR side is the first side the SQZ path encounters. In other words, the LO encounters the AR side first.

 

The optimization of the BS ratio goes in 3 steps:

Step 1:

I first removed the lenses and tuned the ratio of the BS while looking at the DC channels. The objective is to minimize the DC Diff. It will be minimized at a non-zero value (typ 100mV) because the optical paths to each PD are not the same: one beam will have a larger waist than the other so we expect different photocurrents.

Step 2:

I moved the R=25mm SP lenses back by about 1 inch such that the PD are both about 2 inches from the lenses. I expect that the new waist is at the position of the PDs.

Step 3: final tuning of the BS to zero the Diff (typically target +-5mV).

 

Then I took some spectra of the shot noise. I forgot to take the electronic noise.

The input power on the BS is 3.37mW.

Non-image files attached to this report
Comments related to this report
jacques - 16:21, Friday 08 March 2024 (11547)

The shot noise level on these plots is twice as high as the one on the first homodyne (logbook entry https://emvogil-3.mit.edu/lasti/index.php?callRep=11514), which is strange since the LO power has been the same (~3mW on the BS) and the HD PCB boards should be the same. Also the SR floor levels are different. Thus we need to cross check the shot noise on both HDs and see if the difference can be simply explained by the SR having different calibration settings at the times of the measurements.