hudsonl - posted 19:41, Monday 26 February 2024 (11534)
SHG to OPO Mode Matching
The initial mode match from the SHG to the OPO was done using measured ray transfer matrices of the EOM and Faraday. This solution gave respectable mode matching, but the eventual mode matching looked to be no better than 30% or so. I measured the beam after the EOM and Faraday and used this beam to re-compute the mode matching solution. A la mode agrees that the previous mode matching solution should not have exceeded 40% or so.

The new mode matching solution is to have a f = -32mm lens placed 133mm before the PDH board's M4 and a f = 75.6mm lens placed 42mm before the PDH board's M4. With the thin lens approximation and optimized lens locations, this should give 96% mode matching, and with thick lenses placedin the positions optimal for thin lenses, this should give 78% mode matching. As a result, I expect to achieve somewhere between 80 and 95% mode matching of the beam from the SHG to the OPO. The attached file shows this mode matching solution.