Displaying report 1-1 of 1.
Reports until 00:22, Saturday 24 February 2024
cSQZ General
jacques - posted 00:22, Saturday 24 February 2024 (11532)
PZT driver broken again and then fixed by Eric

While trying to take a spectrum today (Friday), I noticed that there was overall extra noise. Then I had the problem again on the PZT Driver suddently not amplyfing neither the potentiometer nor the RP signal, like in entry https://emvogil-3.mit.edu/lasti/index.php?callRep=11526

I also noticed that the CTRL IN was contaminated by a lot of the VCO's 51MHz signal (+-200mV, to be compared to the RP's 0-1V range).

Eric diagnosed the PZT driver board and found that:

1/ Resistor R9 was apparently not soldered properly

2/ The ground power input had an inductance L2 which prevented high frequencies from being grounded

3/ some other resistors and capacitors needed to be added to better isolate the amplification stage from the VCO's 51MHz crosstalk.

4/ The potentionmeter had to be removed as it was causing more trouble (drift) than what it is worth.

After these changes, he tested the VCO on standalone with a signal generator. We saw that it works properly on MON OUT (x100/100 = same output as CTRL IN).

The next step is to recover the PLL with these fixes.

When reshuffling the layout of the PLL, we should separate as much as possible the VCO from the power supplies and the PZT amplifiers, to avoid crosstalk of any kind.