epolini - posted 14:56, Thursday 22 February 2024 (11531)
Green lab computer connected simultaneously to the internet and the Fuyu stage

Fred managed to solve the issue preventing us from connecting the computer simultaneously to the internet and the Fuyu stage by using an USB-ethernet adaptor and by resetting the Fuyu controller using a reset hole on the controller and removing the power for some time.