Reports until 10:51, Thursday 15 February 2024
cSQZ General
hudsonl - posted 10:51, Thursday 15 February 2024 (11517)
EOM Driver Cabling
I moved the high-power RF amplifiers driving the EOMs used for the SHG and OPO PDH locks from the optics table to the top of the CSQZ electronics box. In their new positions, the coaxial cables between the high-powered amplifiers and the EOMs needed to be extended, so I used new cables. The first two of these significantly distorted my PDH error signal (see plots below). This problem seems to be fixed now that I have a BNC cable from the RF amplifier to the EOM instead of a coax cable with a SMA connector on one end and a BNC connector on the other end.

Attached files show the error functions with two different bad coax cavles and with a BNC cable.
Images attached to this report