epolini - posted 18:39, Wednesday 14 February 2024 (11516)
Laser frequency stabilization loop closed

Today I closed the loop for the laser frequency noise stabilization analogically, then we will close it using FPGA. I used the SR560 as servo, the Marconi as VCO, SG3282 as signal generator and a minicircuit mixer. The scheme implemented is shown in Fig. 1. As a filter I am using an integrator with a gain: LP corner frequency = 1kHz and gain = 10.

To measure the OLTF I included a second SR560 set on DC and gain = 1. I sent a swept since between 100Hz-100kHz of amplitude 500mV and the OLTF is shown in Fig.2. We have UGF ~ 10kHz and gain 20dB at low frequency.

In Fig. 3, the error and correction signals closed loop are shown calibrated in rad/rtHz, in blue and orange respectively. The calibration of the error signal is done by looking at the amplitude of the fringes of the interferometer on the oscilloscope which is 50mV. The calibration of the correction signal is done considering the FM deviation which is 10kHz and then rescale with past measurement and it gives 6.5kHz/V and devide the obtained spectrum by the frequency. The gain at low frequency is not enough to correct for all the noise, but we will do a noise budget to get a worst case estimate of the frequency noise.

Moreover, I turned off and on again the laser and I checked again the error and correction spectra afterwards. Right after turning the laser on, the spectra in V/rtHz looks as shown in Fig. 4. The peak at 8kHz is the one driving fringes for the first 3 minutes after the laser is emitting and the noise bump at low frequencies disappears. It comes back after few minutes.

Images attached to this report
Displaying report 1-1 of 1.