Below are noise spectra taken with the Agilent 4395A on the HD that was made by an undergrad and that we took from purple lab. This spectrum analyzer goes higher in frequency than the SR785 but its noise floor is quite high. I did not amplify the signal from the HD.
Notice that the y scale is the power spectrum in dBm/Hz (not the amplitude spectrum in Vrms/sqrt(Hz)) but this does not change the following conclusions of the analysis.
We see clearly that we are no longer shot noise limited from 100kHz upwards. Rather we are limited by the electronics. After discussion with Eric, we should devise and build a new board which has lower effective capacitance.