I rebuilt the HD in a more compact way on a 12in x 18in x 2cm board. I roughly aligned all the waveplates, mirrors and the BS to balance the power incident on the HD. I took the following power budget:
Power budget:
4mW at the output of the LO fiber
3.16mW after the PBS+l/4+PZT setup, going to the 50/50 BS
1.60mW transmitted from the BS
1.57mW reflected from the BS.
When I try to power up the HD and read its PD outputs on an oscilloscope, I only see a non-zero voltage on the port "PD1" (channel 2 of the picture). I don't see anything on the port "PD2" (channel 1, notice the voltage scale) even though Ì can clearly see the IR spot aligned with the PD using a beam card. I don't understand what's happening. Is PD2 = PDA broken?
The solution was to push PD2 well into its socket and then the electrical connection was ok. The goal is then to tune the angle of the BS to simultaneously:
1/ maximimze the DC power on each PD
2/ zero the DC OUT signal (differential)
Next step: The verification of how well the BS is balanced can be seen on a spectrum of DC OUT (see figure attached by Hudson), the sooner the spectrum is shot noise limited the better (here 1000 Hz).