cSQZ General
hudsonl - posted 19:57, Tuesday 06 February 2024 (11500)
SHG to OPO Alignment via PDH Extension Board
I have initial alignment from the SHG to the OPO using the PDH extension board. This alignment was much less painful than the initial OPO alignment, mostly because the OPO itself was already aligned, and the CLF was aligned to the OPO. My alignment procedure was as follows:

- Ensure that the CLF is well aligned to the OPO
- Remove the OPO from the compact squeezer board
- Use pinholes to align the SHG beam to the CLF beam for coarse alignment
- Using the aligned SHG beam, place a 50/50 beam splitter with a lens and a PD on one path and a lens and a CCD on the other path
- Place the OPO back on the compact squeezer board
- Align the retro-reflected 532nm beam from the OPO to itself using a beam card in the near and far fields

At this point, I was able to see cavity flashes on the CCD while scanning the OPO. This allowed me to adjust the alignment of the beam to the OPO. The attached image shows the end point of this alignment procedure.

From looking at the mode scan on the CCD, the remaining discrepancy between the SHG beam and OPO appears to be due to mode-mismatch rather than from misalignment. Tomorrow, I will adjust the positions of the lenses between the SHG and OPO to improve the mode matching.