The PLL has been swapped to the new driver and the loop could be closed. Everything seems to be working as intended so far. Thanks to the increased range of the PZT driver, the loop can be closed during >1h without manual intervention.
I have not computed the phase noise yet, nor have I optimized the loop.
As reference, the loop can be closed with P=0.2, I=20.
Below a screenshot of the error signal on channel 1 (scale 0.5V/div whereas the error saturates at +-2V. So the error signal is not railed).