Displaying report 1-1 of 1.
Reports until 18:43, Friday 02 February 2024
cSQZ General
jacques - posted 18:43, Friday 02 February 2024 - last comment - 18:46, Friday 02 February 2024(11496)
Characteristics of the custom PZT driver for PLL

I compile here the way the custum PLL PZT driver (modified by Eric) works:

- HV IN: power supply (0-100V) connects here

- CTRL IN: RP correction signal connects here

- HV OUT: high-voltage signal which is sent to the laser's PZT. The HV OP amp has a gain of 100 at frequencies below 50Hz.

- MON OUT: a 1/100 signal which mimics the HV OUT port, for monitoring.

There is an extra potentiometer which should be set at 0V for our purposes.

The BNS port between MON OUT and CTRL IN is not connected to anything.

Comments related to this report
jacques - 18:46, Friday 02 February 2024 (11497)

After powering the driver with both a +/-18V and 0/100V power supplies, I sent a slow ramp from teh RP to the driver and looked at the monitor output. I could see that the MON OUT went from 0 to 1V during the scan. I had previously made the mistake of using a 0/18V power supply, instead of a +/-18V. 

Conclusion: we can test finally that the HV OUT is saturated at 100V, then move to the actual actuation of the DBL PZT. 

Displaying report 1-1 of 1.