I have realigned the PLL's optical setup, whose power budget is as follows:
700uW at the start of the AOM path (ie incident on the AOM)
AOM driven using the Moglabs at 34dBm at 51.6MHz
600uW out of AOM, incuding 200mW of frequency-shifted light.
52.4uW from AOM path, measured at PLL PD
170uW from DBL path measured at PLL PD.
I then swapped the AOM driver, from the moglabs, to the VCO. the power on the PLL PD has increased:
75uW from AOM path, measured at PLL PD.
I tried to close the PLL with the RP -> Thorlabs Driver -> DBL EXT but I again have the problem of the error signal not going to zero.
I will not investigate this further because this is not the final setup; rather I will switch the Thorlabs driver to the custom one by Eric and study the loop from there.