Displaying report 1-1 of 1.
Reports until 17:40, Thursday 01 February 2024
cSQZ General
jacques - posted 17:40, Thursday 01 February 2024 - last comment - 18:31, Thursday 01 February 2024(11492)
PLL recovery (1): power budget

I have realigned the PLL's optical setup, whose power budget is as follows:

700uW at the start of the AOM path (ie incident on the AOM)

AOM driven using the Moglabs at 34dBm at 51.6MHz

600uW out of AOM, incuding 200mW of frequency-shifted light.

52.4uW from AOM path, measured at PLL PD

170uW from DBL path measured at PLL PD.

I then swapped the AOM driver, from the moglabs, to the VCO. the power on the PLL PD has increased: 

75uW from AOM path, measured at PLL PD.

 

 

Comments related to this report
jacques - 18:31, Thursday 01 February 2024 (11495)

I tried to close the PLL with the RP -> Thorlabs Driver -> DBL EXT but I again have the problem of the error signal not going to zero. 

I will not investigate this further because this is not the final setup; rather I will switch the Thorlabs driver to the custom one by Eric and study the loop from there.

Displaying report 1-1 of 1.