After moving the compact squeezer and the electronics box, the SHG alignment had changed, although we still saw clear cavity flashes. It was easy to re-align the beam into the SHG by adjusting the tip/tilt of the fiber collimator and the PBS. Attached images show cavity transmission scans prior to and after alignment.