epolini - posted 11:55, Wednesday 10 January 2024 (11467)
Polarization controller installed and power budget

Yesterday, with the help of Fred, we managed to connect the polarization controller POS-002 (general photonics) to the computer. We used a computer power supply at +/-12V since we are running out of other easy solutions. The pinage of the power supply is shown in Fig.1 and how the cables have been connected in Fig.2. The manual of the polarization controller can be found here. The black cable from the POS-002 is the pin 2 and correspond to the +12V pin, the white one to ground and the grey one to -12V. We connected the black cable to pin 10 of the power supply, the white one to pin 17 of the power supply and the grey one to pin 14 of the power supply. In order to turn the power supply on we connected pin 16 to pin 18. Fig. 3 shows how the setup looks like.

Initially, I connected the 9-pin cable from the POS-002 directly to the computer and we tried different software to communicate with it (termite, SecureCRT, YAT). We verified that the controller was powered, that the pinage was correct and that the instrument was seen by the computer. We coudn't make it work. Fred tried to connect the cable using an USB adaptor and then we were able to speak with it using YAT (a software that now is on the computer desktop). The commands are written on the POS-002 manual. I was able to enable and disable the controller via command lines. The next step would be to use the Labview scripts to change the feedback loop. 

I also adjusted the setup to avoid the PD 1811 satutation. 

Power budget (input power 200mW):

Since on the short arm the power is more than double the power on the long arm, and the two are recombining at the second 50/50 BS in front of the PD, I installed a 50/50 PM BS there.

The new power on the short arm is 2.78 mW. The power reaching the 1811 PD is now 2.14 mW.

The signals from the PD were saturating, so I installed a 99/1 PM BS after the second 50/50 BS. The powers at this BS are:

I connected a fiber dump to port 3 and I sent port 4 to the PD. The DC power read on the oscilloscope from the 1811 is 280 mV and the pk-pk voltage of the RF channel is 1.15 V. 

The final setup is shown in Fig. 4.

Images attached to this report