We have replaced the fiber box with connectorized fiber. We send roughly 250mW of power towards the first fiber splitter in the chain, which has a coupling efficiency of roughly 50% as of the last time we measured it. On the SHG pump path, this power then goes through a faraday isolator, a 99/1 beam splitter (1% is sent to the PLL), a 90/10 splitter (10% is sent to the CLF and LO paths), a patch cable, a U-bench with a free-space EOM, polarization paddles, and finally to the fiber to free-space collimator on the compact squeezer optics board. Currently, with 250 mW of power launched at the fiber, we get about 17mW of power out of the SHG's fiber collimator. A large source of loss on this path is the U-bench. Currently, we have about 55mW sent towards the EOM out of the first fiber collimator, but only about 25mW makes it into the fiber with the polarization paddles. We anticipate that the U-bench should be able to be aligned such that it has 70-80% coupling. With the current setup, this would give about 26mW or power out of the SHG's collimator. This 26mW is still low, so there are several other ways to reduce losses. (1) Replace the thorlabs single-mode fiber through the polarization paddles with fiber from ADF, which should be higher quality. We already have this fiber, so this should be easy to try. (2) Remove the patch cable between the 90/10 beam splitter and the U-bench. This patch cable is not necessary in the current configuration. And its inclusion likely introduces about 0.3dB of loss. (3) Splice the fibers together. This should reduce losses through the fiber network, but will revert us back to the scenario where the fiber components are monolithic. We would like to avoid this if possible. (4) Remove the polarization paddles and replace them with ThorLabs U-bench mounted lambda/2 and lambda/4 waveplates after the EOM. These waveplates have less than 1% loss each, so the total coupling from the U-bench input to SHG pump light would be 70-80% efficient, compared to about 50% efficiency currently (even if the U-bench was aligned so it had 70% efficiency). This is easy to implement, but will cost about $700.