jacques - posted 18:25, Thursday 04 January 2024 (11464)
Weird response on SHG MASTER RP's NA

I tested the "SHG" (MASTER) RP's network analyzer function, using the diagram: Excitation -> out1 -> in1 -> NA, where a short SMA cable was used for the connection from out1 to in1. The Excitation is sent directly to out1; there is no PID in the setup. 

Below are figures of what I get for the frequency response, for testing out1->in1 and out2->in2 respectively. 

Interpretation: looking at the magnitudes, there are weird resonances at 15MHz (out2) up to +6dB, and a double peak at 30MHz (out1) up to 20 dB. We are below the Nyquist frequency of 62.5MHz so I don't understand why these resonances exist. The phase's rolloff at >100kHz is probably due to some internal delay to the RP.

Conclusion: I don't investigate this further because this RP is used for the PZT drive only (limited at 50kHz anyways), but we should keep this in mind for future reference. Hudson showed me plots of the same measuresments on another RP where this resonance did not exist.